Performance Data

Within-run reproducibility

[sec]

  Sample 1 Sample 2
1 31.3 59.2
2 31.8 60.3
3 31.7 61.0
4 32.0 60.8
5 31.3 60.8
6 32.1 60.3
7 32.1 60.6
8 31.9 60.6
9 31.5 60.5
10 32.0 60.3
Mean 31.8 60.4
S.D. 0.31 0.50
C.V. (%) 1.00 0.83
Max. 32.1 61.0
Min. 31.3 59.2
Range 0.8 1.8

On-Board Stability (After Reconstitution)

[on Coapresta 2000]

On-Board Stability (After Reconstitution)

Interference

  Concentration Measured Value(sec)
Base Plasma Including Interfering Substance
Unconjugated
Bilirubin
20 mg/dL 28.0 28.5
Conjugated
Bilirubin
20 mg/dL 28.3 28.4
Hemoglobin 500 mg/dL 28.5 28.2
Chyle 3000 FTU 28.3 28.4

LA Sensitivity (Mixing Study)

Using CP2000’s multipoint dilution function, LA positive samples and coagulation factor deficient plasma were measured at 7 points. LA positive sample produces a concave graph while the coagulation factor deficient sample produces a convex graph.

LA Sensitivity (Mixing Study)

Heparin Sensitivity

Heparin Sensitivity

Correlation

Correlation

Assay Procedure

Assay Procedure

Coagpia APTT-N

Additional information

Coagpia APTT-N | APTT Reagent

10×4mL, Storage: 2-8℃

Coagpia APTT-N | Calcium Chloride Solution

10×10mL, Storage: 2-8℃

Control P-N for Coagpia*

5×2conc.×1mL(lyophilized), Storage: 2-8℃

Scroll to top