Nanopia CRP

Monoclonal antibody and specially-made latex allows measurement from low (0.01 mg/dL) to high (42 mg/dL) concentrations.

  • *This package insert follows the Pharmaceuticals, Medical devices and Other Therapeutic Products Act of Japan.

Description

*Read carefully the package insert before use.

Reference Data

Diagram

Diagram

Precision

(mg/dL)

  Sample 1 Sample 2 Sample 3
n 20 20 20
Mean 0.1055 0.2114 8.1719
S.D. 0.0019 0.0028 0.0542
C.V.(%) 1.81 1.31 0.66
Range 0.007 0.01 0.23

Linearity

Linearity

Correlation

Correlation

Prozone

Prozone

Interference

  Additive amount Measurement(mg/dL)
Base serum Added serum
Hemoglobin 500mg/dL 0.37 0.36
Unconjugated Bilirubin 50mg/dL 0.36 0.37
Conjugated Bilirubin 50mg/dL 0.37 0.37
Ascorbic acid 50mg/dL 0.37 0.37
Lipid emulsion 5% 0.36 0.36

Calibration curve

Calibration curve

Functional sensitivity

Functional sensitivity

Detection limit

Detection limit

Nanopia CRP

Additional information

Nanopia CRP |CRP Buffer 1

1×100mL, 2x85mL, 2x50mL, Storage: 2-10℃

Nanopia CRP | CRP Latex Reagent 2

1x100ml, 2x85ml, 2x50ml, Storage: 2-10℃

Scroll to top