Reference Data

Within-run reproducibility

(µg/mL)

  Sample 1 Sample 2
1 3.5 9.5
2 3.6 9.5
3 3.5 9.4
4 3.4 9.3
5 3.5 9.5
6 3.5 9.3
7 3.5 9.4
8 3.6 9.5
9 3.5 9.5
10 3.5 9.3
Mean 3.5 9.4
S.D. 0.06 0.09
C.V. (%) 1.6 1.0
Max. 3.6 9.5
Min. 3.4 9.3
Range 0.2 0.2

On-Board Stability

[on Coapresta 2000]

On-Board Stability

Interference

  Concentration Measured Value (µg/mL)
Base
Plasma
Including
Interfering Sample
Unconjugated
Bilirubin
20 mg/dL 2.70 2.80
Conjugated
Bilirubin
20 mg/dL 2.80 2.80
Hemoglobin 500 mg/dL 2.12 2.14
Chyle 2500 FTU 2.17 2.04

Linearity

Linearity

Assay Procedure

Assay Procedure

Nanopia D-dimer

Additional information

Nanopia D-dimer | D-dimer Buffer Solution 1

1×10mL, Storage: 2-10℃

Nanopia D-dimer | D-dimer Latex Reagent 2

1×10mL, Storage: 2-10℃

D-dimer Calibrator*

1×6conc.×0.5mL(lyophilized), Storage: 2-10℃