Nanopia SF

  • Measures FM (fibrin monomer) and SFMC (soluble fibrin monomer complex. Detects only FM and SF formed by the action of thrombin on fibrinogen. Does not react with fibrinogen degradation products formed by the action of plasmin
  • Specifically reflects hypercoagulable states
  • Plasma (citrated plasma) samples can be used
  • *This package insert follows the Pharmaceuticals, Medical devices and Other Therapeutic Products Act of Japan.

Description

*Read carefully the package insert before use.

Performance Data

Within-run reproducibility

(µg/mL)

  Sample 1 Sample 2
n 10 10
Mean 11.23 51.63
S.D. 0.07 1.35
C.V. (%) 0.64 2.62
Max. 11.31 54.59
Min. 11.09 50.57
Range 0.22 4.02

Interference

Substance added Concentration Measurement (µg/mL)
Base Plasma Added Plasma
Unconjugated Bilirubin 20 mg/dL 23.5 24.0
Conjugated Bilirubin 20 mg/dL 23.3 23.5
Hemoglobin 100 mg/dL 23.0 24.9
Chyle 2000 FTU 24.5 22.9

Assay procedure

Assay procedure

Nanopia SF

Additional information

Nanopia SF | SF Buffer Solution 1

1×10.5mL, Storage: 2-10℃

Nanopia SF | SF Latex Reagent 2

1×10mL, Storage: 2-10℃

SF Calibrator*

1×5conc.×0.5mL(lyophilized), Storage: 2-10℃

SF Control*

3×2conc.×1mL(lyophilized), Storage: 2-10℃