NORUDIA Insulin

  • Ready-to-use liquid reagent
  • Can be used on various automated analyzers
  • Blood sugar and insulin can be measured at the same time from a glucose tube (NaF tube)
  • Good reagent stability in non-calibration.
  • *This package insert follows the Pharmaceuticals, Medical devices and Other Therapeutic Products Act of Japan.

Description

*Read carefully the package insert before use.

Reference Data

Reference values measured by Hitachi 7180

Reproducibility

  Measured Value (µU/mL)
Level 1 Level 2 Level 3
n 10 10 10
Mean 6.8 20.4 120.0
S.D. 0.07 0.09 1.42
CV(%) 0.99 0.47 1.18
Max. 6.9 20.5 122.2
Min. 6.7 20.2 118.0

Detection Limit

Detection Limit

Prozone

Prozone

Correlation

Correlation

Interference

  Added Amount Measured Value (µU/mL)
Base Sample Spiked Sample
Conjugated Bilirubin 20mg/dL 12.7 12.7
Unconjugated Bilirubin 20mg/dL 12.7 12.9
Chyle 2000 FTU 12.6 13.3
Intralipos 5% 12.2 12.0
Rheumatoid Factor 500IU/mL 16.8 18.6
Gastrin 10µg/mL 14.2 14.2
Glucagon 10µg/mL 14.2 13.9
Secretin 10µg/mL 14.2 13.4
Proinsulin 12.2ng/mL 0.0 0.0
C peptide 3.9ng/mL 0.0 0.0

On-board Stability

On-board Stability

Measurement Range (on Hitachi 7180)

1-150µU/mL

Reference Interval

Fasting insulin level: 5-10 µU/mL

NORUDIA Insulin

Additional information

NORUDIA Insulin|Insulin Buffer Solution 1

2×15mL, Storage: 2-10℃

NORUDIA Insulin|Insulin Buffer Solution 2

2×5mL, Storage: 2-10℃

Insulin Calibrator*

1vial×6oncentration for 1mL, Storage: 2-10℃

Insulin Control*

3vials×2concentration for 1mL, Storage: 2-10℃