NORUDIA CYSTATIN C

  • Highly specific reagent using monoclonal antibodies
  • Minimal interference from RF
[Advantages of Measuring Cystatin C]
  • Uses serum or plasma – no complicated urine collection procedures required
  • Minimally affected by age, sex, muscle mass, or cancer (prerenal factors)
  • Can detect renal dysfunction earlier than traditional markers
  • *This package insert follows the Pharmaceuticals, Medical devices and Other Therapeutic Products Act of Japan.

Description

*Read carefully the package insert before use.

Reference Data (on Hitachi 7170)

Assay conditions
Sample: 2.4µL
Cystatin C Buffer Solution 1 120µL
Cystatin C Latex Reagent 2 120µL

Within run reproducibility

  Measured Value (mg/L)
Low Concentration High Concentration
n 10 10
Mean 0.492 3.941
S.D. 0.004 0.045
CV (%) 0.85 1.15
Max. 0.50 4.04
Min. 0.49 3.89

Detection Limit

Detection Limit

Prozone

Prozone

Interference

  Added volume Measured Value (mg/L)
Base sample Added sample
Unconjugated Bilirubin 20mg/dL 0.68 0.68
Conjugated Bilirubin 20mg/dL 0.68 0.68
Chyle 2000 FTU 0.67 0.69
Hemoglobin 500mg/dL 0.68 0.67
Rheumatoid Factor 500IU/mL 0.75 0.75

On-board Stability

On-board Stability

Correlation

Correlation

Measurement Range:

0.1-10mg/L

Reference Interval:

0.5-1.0mg/L

NORUDIA CYSTATIN C

Additional information

NORUDIA CYSTATIN C|CYSTATIN C Buffer Solution 1

2×12mL, Storage: 2-10℃

NORUDIA CYSTATIN C|CYSTATIN C Latex Reagent 2

2×12mL, Storage: 2-10℃

Calibrator for NORUDIA CYSTATIN C*

5conc.×2mL, Storage: 2-10℃

Control for NORUDIA CYSTATIN C*

3×2conc.×2mL, Storage: 2-10℃